Related Experiment Video
Updated: Oct 1, 2025

Lab-Scale Model to Evaluate Odor and Gas Concentrations Emitted by Deep Bedded Pack Manure
Published on: July 19, 2018
Quantitative evaluation of chemical fume hoods performance by CO2 tracer gas
Farshid Ghorbani-Shahna1, Maryam Feiz-Arefi2,3
1Center of Excellence for Occupational Health, Occupational Health and Safety Research Center, School of Public Health, Hamadan University of Medical Sciences, Hamadan, Iran.
Background:
Various chemical substances and carcinogens have been presented in medical sciences universities' educational and research laboratories. For this purpose a suitable ventilation system had to be implemented to ensure the correct operation of the hoods.
Objective:
To evaluate the performance of laboratory chemical fume hoods of the University of Medical Sciences using a novel quantitative method.
Methods:
In this study, 43 chemical fume hoods were investigated in the laboratories of the University of Medical Sciences. The technical specifications of the hoods and their compliance with the standard have been investigated. The hoods face velocity was measured using a thermal anemometer. Quantitative evaluation was performed using the new method of CO2 tracer gas and the results were analyzed using SPSS software version 19.
Findings:
The hoods presented both favorable and unfavorable results in terms of technical specifications and location. The results showed 50.2% of the hoods have visible leakage. Hood face velocity was not suitable for any of the hoods in the case fully open.when half open only 16.3% of the hoods and in the case of 25% open face, 34.9% of the hoods had a good velocity. Most hoods have CO2 leakage even at small amounts.
Conclusions:
the unsuitable performance of the hoods is mainly due to the unsuitability of the fans, furthermore investigation and correction of technical problems are required. The new quantitative method is a suitable method for routine evaluating chemical fume hoods and can replace the SF6 gas tracer method.
More Related Videos
07:24Combustion Chemistry of Fuels: Quantitative Speciation Data Obtained from an Atmospheric High-temperature Flow Reactor with Coupled Molecular-beam Mass Spectrometer
Published on: February 19, 2018
08:18Design and Use of a Full Flow Sampling System FFS for the Quantification of Methane Emissions
Published on: June 12, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...