Staged correction trajectory with hexapod external fixator for the satisfactory reduction of long bone shaft fracture
Yanshi Liu1, Fei Wang2, Kai Liu1
1Department of Trauma and Microreconstructive surgery, the First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
Backgroud:
When the reduction of long bone shaft fracture fragments is performed by a hexapod external fixator, the collision and interference between bony ends often results in an incomplete reduction and a time-consuming procedure. The purpose of this study was to present and determine the clinical effectiveness of staged correction trajectory with hexapod external fixator in the reduction of a long bone shaft fracture.
Methods:
A total of 57 patients with tibial shaft fractures treated by hexapod external fixator were retrospectively analyzed from June 2016 to February 2020. Thirty-one cases (Group I) underwent a conventional one-step reduction trajectory from June 2016 to July 2018. Starting in September 2018, the other twenty-six patients (Group II) underwent staged correction trajectory (three key points reduction trajectory of "distraction-derotation-reduction"). The demographic data, residual deformities before and after correction, number of repeated X-rays after the first postoperative X-ray, duration of deformity correction process, and external fixation time were analyzed. Johner-Wruhs criteria were used to evaluate the final clinical outcomes.
Results:
All the 57 patients achieved satisfactory fracture reduction and bone union. There were no significant differences between the two groups in demographic data, residual deformities before and after correction, external fixation time, and final clinical outcomes (p > 0.05). The average number of repeated X-rays after the first postoperative X-ray and mean duration of deformity correction process in Group II (1.3 times, 2.9 days) were all less than those in Group I (2.3 times, 5.1 days) (p < 0.05).
Conclusion:
Compared with the conventional one-step reduction trajectory, there is no differences in final clinical outcomes, but the staged correction trajectory provides less repeated X-rays and shorter reduction process duration.