Comparison of the sporicidal activity of a UV disinfection process with three FDA cleared sterilants
Lionel Pineau1, Caroline Radix1, David J Weber2
1Division of Medical Device Testing, Eurofins Biotech Germande, Aix-en-Provence, France.
Background:
Endocavitary probes are semi-critical devices and must undergo, at least, high level disinfection (HLD) between uses. Therefore, they should be high level disinfected between uses (i.e., with a product/process that kills all forms of microbial life; bacteria, fungi, mycobacteria, and virus, and in some countries, a demonstrated potential for sporicidal activity). In this study, the sporicidal activity of three common Food and Drug Administration cleared sterilants (CIDEX OPA Solution, SPOROX II Sterilizing and Disinfection Solution and CIDEX Activated Dialdehyde Solution) was compared with the sporicidal activity of an ultraviolet disinfection technology (Hypernova Chronos, Germitec) against Bacillus subtilis ATCC 19659 spores spread on silicone flat carriers in the presence of inorganic and organic soil.
Results:
The results indicate that the UV disinfection process presented within a 35 seconds exposure time a sporicidal efficacy substantially higher than the chemical sterilants used according to manufacturer instructions for HLD.
Conclusions:
This study demonstrated that even if it cannot be tested/approved as a sterilant according to AOAC 966.04, the UV unit is much more effective than usual Food and Drug Administration approved chemical HLD products to kill spores in real use conditions. This finding questions the relevancy of evaluating product efficacy within extended conditions giving results that could mislead users to select the most effective HLD product/process for the reprocessing of their medical devices.
More Related Videos
14:17The Portable Chemical Sterilizer PCS, D-FENS, and D-FEND ALL: Novel Chlorine Dioxide Decontamination Technologies for the Military
Published on: June 29, 2014
10:03Investigating the Detrimental Effects of Low Pressure Plasma Sterilization on the Survival of Bacillus subtilis Spores Using Live Cell Microscopy
Published on: November 30, 2017
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Chemical Agents for Microbial Control
Methods for Controlling Microbial Growth