Related Experiment Video
Updated: Sep 30, 2025

An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Evaluation of Triggered Electromyogram Monitoring during Insertion of Percutaneous Pedicle Screws
Hayato Futakawa1,2, Shigeharu Nogami1, Shoji Seki2
1Department of Orthopedic Surgery, Takaoka City Hospital, Toyama 933-8550, Japan.
Objective:
percutaneous pedicle screw (PPS) fixation has been widely used in minimally invasive spine stabilization. Triggered electromyogram (TrEMG) monitoring is performed to prevent PPS misplacement, but is not widely accepted. We have newly developed an insulating tap device to minimize the misplacement of PPS.
Methods:
TrEMG was measurable in insulation tap devices in 31 cases, and in non-insulating tap devices in 27 cases. Fluoroscopy was used to insert 194 PPS and 154 PPS, respectively. Based on the Rampersaud classification of postoperative computed tomography, we classified PPS insertion into four categories (Grade A as no violation, Grade D as more than 4 mm perforation).
Results:
Grade A was noted in 168 PPSs (86.6%) and Grade B to D in 26 PPSs in the insulation tap device group, and Grade A was noted in 129 PPSs (83.8%) and Grade B to D in 25 PPSs in the non-insulating tap device group, respectively. At a cutoff value of 11 mA, the sensitivity was 41.4% and the specificity was 98.2%. The sensitivity and specificity of the non-insulating tap device were 4.0% and 99.2%, respectively.
Conclusions:
The insulation treatment of the tap device has improved the sensitivity of TrEMG. TrEMG using the insulating tap device is one of the methods for safe PPS insertion.