Dosimetry of a novel focused monoenergetic beam for radiotherapy
Jamshid Moradi-Kurdestany1, Dirk A Bartkoski1, Ramesh Tailor1
1Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, United States of America.
Abstract:
Objective. A novel treatment modality is currently being developed that produces converging monoenergetic x-rays. Conventional application of dosimetric calibration as presented in protocol TG61 is not applicable. Furthermore, the dosimetry of the focal point of the converging beam is on the order of a few millimeters, requiring a high-resolution dosimeter. Here we present a procedure to calibrate radiochromic film for narrow-beam monoenergetic 60 keV photons as well as absolute dosimetry of monoenergetic focused x-rays. A study of the focal spot dose rate after passing through a bone-equivalent material was also done to quantify the effects of heterogeneous materials.Approach.This was accomplished by configuring a polyenergetic beam of equivalent energy using a clinical orthovoltage machine. Calibrated films were then used to perform absolute dosimetry of the converging beam by measuring the beam profile at various depths in water. Main Results.A method for calibrating radiochromic film has been developed and detailed that allows absolute dosimetry of a monoenergetic photon beam. Absolute dosimetry of a focused, mono-energetic beam resulted in a focal spot dose rate of ∼30 cGy min-1at a depth of 5 cm in water.Significance.This work serves to establish a dosimetry protocol for mono-energetic beam absolute dosimetry as well as the use of such a method for measurement of a novel teletherapy modality.
More Related Videos
Related Concept Videos
Radiation: Applications
The average...
Biological Effects of Radiation
Absorption of Radiation
Isotopes and Radioisotopes
An isotope containing...