A prospective evaluation of diagnostic performance of a combo rapid antigen test QuickNavi-Flu+COVID19 Ag
Yuto Takeuchi1, Yusaku Akashi2, Yoshihiko Kiyasu3
1Department of Infectious Diseases, University of Tsukuba Hospital, 2-1-1 Amakubo, Tsukuba, Ibaraki, 305-8576, Japan; Division of Infectious Diseases, Department of Medicine, Tsukuba Medical Center Hospital, 1-3-1 Amakubo Tsukuba, Ibaraki, 305-8558, Japan.
Introduction:
Since respiratory sample collection is an uncomfortable experience, simultaneous detection of pathogens with a single swab is preferable. We prospectively evaluated the clinical performance of a newly developed antigen test QuickNavi-Flu+COVID19 Ag (Denka Co., Ltd., Tokyo, Japan) which can detect severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and influenza viruses at the same time with a single testing device.
Methods:
We included those who were suspected of contracting coronavirus disease 2019 (COVID-19) and were referred to a PCR center at Ibaraki prefecture in Japan, between August 2, 2021 to September 13, 2021, when the variant carrying L452R spike mutation of SARS-CoV-2 were prevalent. Additional nasopharyngeal samples and anterior nasal samples were obtained for the antigen test and were compared with a reference real-time reverse transcription PCR (RT-PCR) using nasopharyngeal samples.
Results:
In total, 1510 nasopharyngeal samples and 862 anterior nasal samples were evaluated. During the study period, influenza viruses were not detected by QuickNavi-Flu+COVID19 Ag and reference real-time RT-PCR. For SARS-CoV-2 detection in nasopharyngeal samples, the sensitivity and specificity of the antigen test were 80.9% and 99.8%, respectively. The sensitivity and specificity using anterior nasal samples were 67.8% and 100%, respectively. In symptomatic cases, the sensitivities increased to 88.3% with nasopharyngeal samples and 73.7% with anterior nasal samples. There were three cases of discrepant results between the antigen test and the real-time RT-PCR. All of them were positive with the antigen test but negative with the real-time RT-PCR in SARS-CoV-2 detection.
Conclusion:
A combo kit, QuickNavi-Flu+COVID19 Ag, showed an acceptable sensitivity and sufficient specificity for SARS-CoV-2 detection, especially using nasopharyngeal sample collected from symptomatic patients.