A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Measuring intracompartmental pressures for the chronic exertional compartment syndrome: Challenging commercially
S Vogels1, E D Ritchie2, E W P Bakker3
1Department of Surgery, Alrijne Hospital, Leiderdorp, the Netherlands; Trauma Research Unit, Department of Trauma Surgery, Erasmus Medical Center, Rotterdam, the Netherlands.
Abstract:
The aim of this study was to compare commercially available manometers and needles used for intracompartmental pressure measurements for accuracy. An experimental compartment simulation model was developed in order to compare four different terminal devices (Compass manometer, Stryker device, Meritrans transducer, and arterial line) and 22 types of needles. First, all possible device/needle combinations were introduced in rubber ports at the bottom of a water column. The water column was gradually drained three times for each device/needle combination and pressures were recorded by two researchers. This procedure was repeated after placement of a sample of homogenous porcine gluteal muscle tissue. When measuring the fluid column only, all Intraclass Correlation Coefficients were found to be ≥0.980, indicating good resemblance to the reference pressure. After addition of the muscle tissue to the experimental set up, accuracy of the Stryker device and arterial line remained the same. The accuracy of the Compass manometer and Meritrans transducer became less. Excellent reliability was found for 7 out of 22 needles when combined with the Stryker device and for 2 out of 22 needles when combined with the arterial line. The four terminal devices tested were all accurate when measuring pressure in a water column, whereas only the Stryker device and arterial line accurately represented the water column pressure in a porcine gluteal muscle model. Future research will have to verify whether these results can be repeated in human cadaver models and in vivo.
More Related Videos
14:52Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
Published on: December 11, 2013
09:37Evaluation of Respiratory Muscle Activation Using Respiratory Motor Control Assessment RMCA in Individuals with Chronic Spinal Cord Injury
Published on: July 19, 2013
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure in the Leg
Preparation:
Measurement of Blood Pressure