[Comparison of the measurement of lactate at the cord by hand-held devices approved for use during labor with a
E Mosnino1, A-S Michel1, E Messedi2
1Service de gynécologie obstétrique, hôpital Trousseau, 26, avenue du Dr-Arnold-Netter, 75012 Paris, France; Sorbonne Université, AP-HP, 91, boulevard de l'Hôpital, Paris, France.
Objective:
To compare two portable lactate devices, Lactate Scout and StatStrip Xpress, to a laboratory reference device by assessing the concordance of cord blood lactate values at birth.
Methods:
We conducted a single-center prospective validation study in a level 3 maternity hospital. For all cord samples taken at birth, we analyzed lactates on the reference device (GEM4000®) and on two portable devices Lactate Scout and StatStrip Xpress. We compared the lactate values from each of the two handheld devices to the reference device.
Results:
A total of 101 samples were collected. Each was analyzed by all three machines. The average lactate value obtained by the different machines was 3.7mmol/L. The mean difference between the Lactate Scout and GEM4000® apparatus was 0.0mmol/L±0.8mmol/L. Between these two devices at lactate threshold of 4.8mmol/L, the negative predictive value was 96,2% (76/79) and the area under the curve was 0.98 (95% CI 0.96-1]. The mean difference between StatStrip Xpress and GEM4000® was 0.1mmol/L±0.7mmol/L. Between these two devices at the lactate threshold of 4.8mmol/L, the negative predictive value was 97,4% (76/78) and the area under the curve was 0.95 (95% CI 0.86-1).
Conclusion:
There is a good correlation between the lactates obtained by the reference device and the two portable devices Lactate Scout and StatStrip Xpress.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement