Related Experiment Video
Updated: Sep 30, 2025

Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
Comparison of point shear wave elastography and transient elastography in the evaluation of patients with NAFLD
Giulio Argalia1, Claudio Ventura2, Niccolò Tosi2
1Departement of Radiological Sciences, Ospedali Riuniti Ancona, Università Politecnica Delle Marche, University Hospital, Via Tronto 10, 60126, Ancona, Italy. giulio.argalia@gmail.com.
Purpose:
To compare point shear wave elastography (pSWE, ElastPQ®) and transient elastography (TE) with Liver Biopsy in order to evaluate fibrosis stage in non-alcoholic fatty liver disease (NAFLD).
Methods:
Our prospective study from September 2017 to October 2020 included 50 consecutively enrolled patients with NAFLD (52.2 ± 13.0 years, 32 male). All patients underwent clinical evaluation, B-mode ultrasound, pSWE, TE and liver biopsy in a single evaluation. The clinical, laboratory and liver biopsy data were compared with liver stiffness (LS) measurement obtained with pSWE and TE. TE and pSWE diagnostic accuracy for the diagnosis of the different fibrosis stages were evaluated using the area under receiver operating characteristic curve (AUROC).
Results:
Only fibrosis stage was independently associated with TE and pSWE. The median liver stiffness measurement for fibrosis stages F0, F1, F2, F3, and F4 using TE was 4.8 (4.7-6.1) kPa, 5.5 (4.4-7.3) kPa, 7.7 (6.1-9.1) kPa, 9.9 (8.8-13.8) kPa, and 20.2 kPa, respectively. The corresponding median liver stiffness measurement using pSWE was 4.2 (4.0-4.8) kPa, 4.7 (4.2-5.8) kPa, 5.1 (4.1-6.9) kPa, 8.5 (5.2-13.3), and 15.1 kPa, respectively. The AUROC of TE for diagnosis of fibrosis stage F1, ≥ F2, ≥ F3, and F4 were 0.795, 0.867, 0.927, and 0.990, respectively. The corresponding AUROC of pSWE was 0.717, 0.733, 0.908, and 1.000, respectively. No association was observed with other histological parameters.
Conclusion:
TE was significantly better than pSWE for the diagnosis of fibrosis stage ≥ F2. No statistically significant differences were found between TE and pSWE AUROC of fibrosis stage ≥ F1, ≥ F3, and F4.
More Related Videos
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
06:09Author Spotlight: Advancing Hepatic Fibrosis Diagnosis Using Magnetic Resonance Elastography and AI
Published on: July 21, 2023