Effect of Mask Selection on the Leak Test in Ventilators Designed for Noninvasive Ventilation
Wei Tan1, Hai-Jia Hou1, Chang-Ling Lu1
1Department of Respiratory and Critical Care Medicine, The First Affiliated Hospital of China Medical University, Shenyang, China.
Background:
The mask leak test used for modern noninvasive ventilators can detect the leak characteristics of masks that are not recommended by the manufacturer, but it has not yet been determined whether this method is acceptable.
Methods:
A noninvasive ventilator equipped with a single-limb circuit and an oronasal mask was connected to a lung simulator. The ventilator was set to S/T mode, and inspiratory positive airway pressure/expiratory positive airway pressure was set to 10/5, 15/5, and 20/5 cm H2O, respectively. Eight nonmanufacturer-recommended oronasal masks were connected to the ventilator. The lung simulator was used to simulate COPD, restrictive disease, and normal lung, respectively. When switching between masks, the mask leak test was set to "Cancel" or "Start Test" in the noninvasive ventilator. The parameters displayed on the lung simulator and ventilator were recorded before and after the mask leak test.
Results:
There were no significant difference before versus after the mask leak test for any lung simulator parameter, including trigger performance (ie, time from the beginning of the simulated inspiratory effort to the lowest value of airway pressure needed to trigger the ventilator, the magnitude of airway pressure drop needed to trigger, and time to trigger), inspiratory pressure delivery, PEEP, tidal volume, and displayed peak inspiratory pressure (all differences < 10%). At different noninvasive ventilation settings, tidal volumes displayed on the ventilator of the 3 masks were significantly different before and after mask leak test (all P < .05, and difference rate > 10%).
Conclusions:
The mask leak test had no effect on the ventilator performance when masks not recommended by the manufacturer were used, but tidal volume monitoring may be more accurate when some masks were used.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...