Application of Biochip Microfluidic Technology to Detect Serum Allergen-specific Immunoglobulin E sIgE
Published on: April 21, 2019
16.5K
[THE CORRELATION OF ALLERGY SPECIFIC IGE MEASUREMENT REAGENT IMMNOCAP, ALASTAT, AND ORITON IGE]
Yasuhiro Yoshikawa1, Tamao Nakashita2, Shinji Motojima2
1Department of Clinical Laboratory, Kameda Medical Center.
Arerugi = [Allergy]
|March 17, 2022
Summary
This study compared Oriton IgE, ImmnoCAP, and Alastat for specific IgE antibody testing. Results show consistent correlations but vary by assay and allergen, with Oriton IgE performing similarly with reduced sample volumes.
Area of Science:
- Clinical Immunology
- Allergen-Specific IgE Testing
Background:
- Limited comparative studies exist for single allergen-specific IgE reagents in Japan, including Oriton IgE, ImmnoCAP, and Alastat.
- Understanding the correlations between these available diagnostic tools is crucial for accurate allergy diagnosis.
Purpose of the Study:
- To investigate the correlations between Oriton IgE, ImmnoCAP, and Alastat for specific IgE antibody testing.
- To evaluate the performance of Oriton IgE with reduced sample volumes.
Main Methods:
- Six common allergens (dust mite, cedar pollen, dog dander, egg white, milk, Candida) were tested using three methods.
- Spearman rank correlation and class-judged agreement were calculated.
- Oriton IgE was further evaluated using small sample volumes.
Main Results:
- Consistent correlations were observed between Oriton IgE and ImmnoCAP/Alastat.
- Correlation patterns differed based on the assay method and specific allergen.
- Oriton IgE showed no significant performance difference between standard and small sample volumes.
Conclusions:
- Comparative data aids in understanding the characteristics of each specific IgE assay.
- Selection of an optimal test method can be informed by these findings.
- Tests requiring smaller sample volumes may be advantageous for pediatric patients.
Related Concept Videos
Cross-reactivity
31.6K
Overview
31.6K
Enzyme-Linked Immunosorbent Assay
14.6K
In 1971, Peter Perlman and Eva Engvall developed an Enzyme-linked immunosorbent assay (ELISA or EIA). ELISA differs from western blot in that the assays are conducted in microtiter plates or in vivo rather than on an absorbent membrane.
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
14.6K