Optimization of the Parameters for the Test Bolus Technique in Computed Tomography Angiography of Head and Neck
1From the Department of Radiology, Hunan Provincial People's Hospital (The First Affiliated Hospital of Hunan Normal University), Changsha, People's Republic of China.
Objective:
The aim of the study was to optimize low-dose scanning protocols for the test bolus technique in 64-detector computed tomography (CT) angiography (CTA) of the head and neck.
Methods:
A total of 250 patients were randomly divided into 5 groups of 50 patients each: conventional group (group A: 120 kV, 100 mA) and low-dose groups (group B: 100 kV, 100 mA; group C: 100 kV, 50 mA; group D: 80 kV, 100 mA; and group E: 80 kV, 50 mA). Subjective scores and objective measurements (CT values of the aortic arch, image noise, and signal-to-noise ratio) were used to evaluate the image quality and compare the radiation doses of the 5 groups.
Results:
The image quality of the 5 groups met the diagnostic requirements, with no significant difference in the subjective evaluation findings (P > 0.05). There were significant differences in the CT values between the conventional group and low-dose groups (P < 0.05); however, there was no significant difference between groups B and C and between groups D and E (P > 0.05). Moreover, the volume CT dose index and dose length product of groups B, C, D, and E decreased by 37.62%, 70.45%, 65.28%, and 83.39%, respectively, compared with those of group A. Although the image noise of the low-dose groups increased, an appropriate reduction in the tube voltage enhanced the contrast medium-induced x-ray attenuation and increased the CT value, which resulted in a nonsignificant difference in the signal-to-noise ratio. Therefore, the image quality of the low-dose groups was not affected compared with that of the conventional group.
Conclusions:
In the test bolus technique in 64-detector CTA, the low-dose protocol of using 80 kV and 50 mA reduces the radiation dose by 83.39% compared with the conventional scan, without affecting the image quality.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...