Finite element analysis to predict short and medium-term performance of the anatomical Comprehensive® Total Shoulder System
Related Experiment Video
Updated: Sep 29, 2025

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Margarida Bola1, José Simões2, António Ramos1
1TEMA, Biomechanics Research Group, Department of Mechanical Engineering, University of Aveiro, Campo Universitário de Santiago, Aveiro 3810-193, Portugal.
Background:
The number of Total Shoulder Arthroplasties (TSA) has increased in these last years with significant increase of clinical success. However, glenoid component loosening remains the most common cause of failure.
Objective:
In this study we evaluated the critical conditions to predict short and medium-term performance of the uncemented anatomical Comprehensive® Total Shoulder System using a finite element model that was validated experimentally.
Methods:
The finite element models of an implanted shoulder analysed included total shoulder components with pegs. The models were simulated in 3 phases of adduction: 45°, 60° and 90° to determine the most critical situation. Two different bone-implant fixation conditions were considered: post-surgery and medium term (2 years).
Results:
These show that the critical condition is for the shoulder in 90° adduction were the highest contact stress (70 MPa) was observed in the glenoid component. Relatively to the interface implant-bone strains, the maximum (-16000 µε) was observed for the short-term in the lateral region of the humerus. The highest micromotions were observed in the central fixation post of the glenoid component, ranging from 20 to 25 µm, and 325 µm in the lateral plane of the humeral component.
Conclusion:
The predicted results are in accordance with clinical studies published and micromotions of the humeral component can be used to predict loosening and to differentiate shoulder implant designs.