Related Experiment Video
Updated: Sep 29, 2025

Author Spotlight: Advancing Biopsy Techniques with Transesophageal Ultrasound
Published on: November 21, 2023
Prospective Study Assessing Impact of Ethylene Oxide Sterilization on Endoscopic Ultrasound Image Quality
Shounak Majumder1, Zaiyang Long2, Alexander W Hooke3
1Department of Gastroenterology, Mayo Clinic, Rochester, Minnesota.
Background & Aims:
Duodenoscope-associated transmission of infections has raised questions about efficacy of endoscope reprocessing using high-level disinfection (HLD). Although ethylene oxide (ETO) gas sterilization is effective in eradicating microbes, the impact of ETO on endoscopic ultrasound (EUS) imaging equipment remains unknown. In this study, we aimed to compare the changes in EUS image quality associated with HLD vs HLD followed by ETO sterilization.
Methods:
Four new EUS instruments were assigned to 2 groups: Group 1 (HLD) and Group 2 (HLD + ETO). The echoendoscopes were assessed at baseline, monthly for 6 months, and once every 3 to 4 months thereafter, for a total of 12 time points. At each time point, review of EUS video and still image quality was performed by an expert panel of reviewers along with phantom-based objective testing. Linear mixed effects models were used to assess whether the modality of reprocessing impacted image and video quality.
Results:
For clinical testing, mixed linear models showed minimal quantitative differences in linear analog score (P = .04; estimated change, 3.12; scale, 0-100) and overall image quality value (P = .007; estimated change, -0.12; scale, 1-5) favoring ETO but not for rank value (P = .06). On phantom testing, maximum depth of penetration was lower for ETO endoscopes (P < .001; change in depth, 0.49 cm).
Conclusions:
In this prospective study, expert review and phantom-based testing demonstrated minimal differences in image quality between echoendoscopes reprocessed using HLD vs ETO + HLD over 2 years of clinical use. Further studies are warranted to assess the long-term clinical impact of these findings. In the interim, these results support use of ETO sterilization of EUS instruments if deemed clinically necessary.
More Related Videos
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Ultrasonography