Evaluation of Corvis ST tonometer with the updated software in glaucoma practice
Ioannis Halkiadakis1, Vasilios Tzimis1, Alexandros Gryparis2
1Ophthalmiatrion Athinon, Athens Eye Hospital, El Venizelou 26 & Sina 2, Athens 10672, Greece.
The updated Corvis ST tonometer shows good agreement with Goldmann applanation tonometry for intraocular pressure and optical-based pachymetry. Biomechanically corrected IOP is not affected by central corneal thickness, offering an advantage for varied corneal measurements.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Corneal Biomechanics
Background:
- Intraocular pressure (IOP) measurement is crucial for glaucoma diagnosis and management.
- Accurate IOP assessment can be influenced by central corneal thickness (CCT).
- The Corvis ST tonometer offers biomechanically corrected IOP (b-IOP) and corneal deformation parameters.
Purpose of the Study:
- To evaluate the agreement between the updated Corvis ST tonometer and traditional methods (Goldmann applanation tonometry - GAT, optical-based corneal pachymetry - OB-CCT).
- To assess b-IOP and Corvis-derived CCT (Corvis-CCT) against GAT IOP and OB-CCT.
- To examine differences in Corvis ST corneal deformation parameters among controls, ocular hypertension (OHT), and primary open-angle glaucoma (POAG) patients.
Main Methods:
- Prospective study including 80 eyes from 80 participants (POAG, OHT, and controls).
- Measurements included GAT IOP, OB-CCT, and updated Corvis ST tonometer readings.
- Bland-Altman analysis was used to assess agreement between measurement methods.
Main Results:
- Good correlation was found between GAT IOP and b-IOP (Spearman's rho=0.767, P<0.001).
- b-IOP showed no significant correlation with OB-CCT (Spearman's rho=-0.13, P=0.917).
- Corvis-CCT correlated well with OB-CCT (Spearman's rho=0.900, P<0.001).
- POAG eyes exhibited a significantly reduced applanation time 2 compared to controls after adjustment.
Conclusions:
- The updated Corvis ST tonometer provides b-IOP and CCT measurements that correlate well with GAT IOP and OB-CCT.
- b-IOP's independence from CCT is advantageous for managing thick or thin corneas.
- Altered corneal viscoelasticity in POAG patients may be reflected by shorter applanation times measured by the Corvis ST.