Parametric image generation with the uEXPLORER total-body PET/CT system through deep learning
Zhenxing Huang1,2, Yaping Wu3, Fangfang Fu3
1Lauterbur Research Center for Biomedical Imaging, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, 518055, China.
Purpose:
Total-body dynamic positron emission tomography/computed tomography (PET/CT) provides much sensitivity for clinical imaging and research, bringing new opportunities and challenges regarding the generation of total-body parametric images. This study investigated parametric [Formula: see text] images directly generated from static PET images without an image-derived input function on a 2-m total-body PET/CT scanner (uEXPLORER) using a deep learning model to significantly reduce the dynamic scanning time and improve patient comfort.
Methods:
[Formula: see text]F-Fluorodeoxyglucose ([Formula: see text]F-FDG) 2-m total-body PET/CT image pairs were acquired for 200 patients (scanned once) with two protocols: one parametric PET image (60 min, 0[Formula: see text]60 min) and one static PET image (10 min, range of 50[Formula: see text]60 min). A deep learning model was implemented to predict parametric [Formula: see text] images from the static PET images. Evaluation metrics, including the peak signal-to-noise ratio (PSNR), structural similarity index measure (SSIM), and normalized mean square error (NMSE), were calculated for a 10-fold cross-validation assessment. Moreover, image quality was assessed by two nuclear medicine physicians in terms of clinical readings.
Results:
The synthetic parametric PET images were qualitatively and quantitatively consistent with the reference images. In particular, the global mean SSIM between the synthetic and reference parametric [Formula: see text] images exceeded 0.9 across all test patients. On the other hand, the overall subjective quality of the synthetic parametric PET images was 4.00 ± 0.45 (the highest possible rating is 5) according to the two expert nuclear medicine physicians.
Conclusion:
The findings illustrated the feasibility of the proposed technique and its potential to reduce the required scanning duration for 2-m total-body dynamic PET/CT systems. Moreover, this study explored the potential of direct parametric image generation with uEXPLORER. Deep learning technologies may output high-quality synthetic parametric images, and the validation of clinical applications and the interpretability of network models still need further research in future works.