Interoperator reliability of an on-site machine learning-based prototype to estimate CT angiography-derived
Yushui Han1, Ahmed Ibrahim Ahmed1, Chris Schwemmer2
1Debakey Heart & Vascular Center, Houston Methodist Hospital, Houston, Texas, USA.
Background:
Advances in CT and machine learning have enabled on-site non-invasive assessment of fractional flow reserve (FFRCT).
Purpose:
To assess the interoperator and intraoperator variability of coronary CT angiography-derived FFRCT using a machine learning-based postprocessing prototype.
Materials And Methods:
We included 60 symptomatic patients who underwent coronary CT angiography. FFRCT was calculated by two independent operators after training using a machine learning-based on-site prototype. FFRCT was measured 1 cm distal to the coronary plaque or in the middle of the segments if no coronary lesions were present. Intraclass correlation coefficient (ICC) and Bland-Altman analysis were used to evaluate interoperator variability effect in FFRCT estimates. Sensitivity analysis was done by cardiac risk factors, degree of stenosis and image quality.
Results:
A total of 535 coronary segments in 60 patients were assessed. The overall ICC was 0.986 per patient (95% CI 0.977 to 0.992) and 0.972 per segment (95% CI 0.967 to 0.977). The absolute mean difference in FFRCT estimates was 0.012 per patient (95% CI for limits of agreement: -0.035 to 0.039) and 0.02 per segment (95% CI for limits of agreement: -0.077 to 0.080). Tight limits of agreement were seen on Bland-Altman analysis. Distal segments had greater variability compared with proximal/mid segments (absolute mean difference 0.011 vs 0.025, p<0.001). Results were similar on sensitivity analysis.
Conclusion:
A high degree of interoperator and intraoperator reproducibility can be achieved by on-site machine learning-based FFRCT assessment. Future research is required to evaluate the physiological relevance and prognostic value of FFRCT.