Purpose:
Self-adjustment of hearing aid amplification enables wearers to customize the hearing aid output to match their preferences and could become an important tool for programming direct-to-consumer devices for people with mild-to-moderate hearing loss. One risk is that user-selected settings may provide inadequate audibility. This study assessed that risk by quantifying relationships between self-adjusted settings, subjective preferences, and speech recognition performance using speech at low levels in quiet, where achieving high speech audibility requires sufficient amplification.
Method:
Fifteen people with symmetric, mild-to-moderate sensorineural hearing loss self-adjusted hearing aid amplification while listening to speech in quiet at 45, 55, and 65 dBA. After self-adjustment, 11 participants made blinded ratings of their self-adjusted fit, their NAL-NL2 prescriptive fit, and experimenter-created fits with reduced gain. Participants completed blinded paired comparisons and sentence recognition assessments using these settings.
Results:
The gain of self-adjusted fits showed a large range of variability between participants. On average, self-adjusted gain was similar to NAL-NL2 prescribed gain for input signals of 55 dBA and slightly greater than prescribed gain for 45-dBA signals. Speech recognition scores for NAL-NL2 fits were consistently high, and differences in speech recognition results were strongly correlated with the overall preferences obtained from paired comparisons.
Conclusions:
Self-adjusted fits are highly variable between individuals for low-audibility conditions. Nonetheless, self-adjusted fits are at least as satisfactory as NAL-NL2 fits, and listeners tend to disfavor settings that result in poorer speech recognition. The findings argue against concerns that self-adjustment will result in inadequate audibility compared to prescribed settings.
More Related Videos
04:32Sound Source Localization Testing in Single-sided Deafness Following Bone Conduction Intervention
Published on: December 20, 2024
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and...
Perceiving Loudness, Pitch, and Location
Place theory, or place coding, suggests that different pitches are heard because various sound waves activate specific locations along the cochlea's basilar membrane. The brain determines the pitch of a sound by...