Detection of frequent neutrophil misclassification by the ProCyte Dx in sick dogs and how to avoid it

E Bergstrand1, H W Tvedten2, I Lilliehöök3

  • 1University Animal Hospital, Swedish University of Agricultural Sciences, Box 7054, 75007, Uppsala, Sweden.

Abstract

Insights

The IDEXX ProCyte Dx analyser commonly misidentifies neutrophils in canine blood, classifying them as monocytes or lymphocytes. Evaluating white blood cell dot plots is crucial for accurate canine neutrophil counts.

Area of Science:

  • Veterinary Hematology
  • Clinical Pathology
  • Diagnostic Cytometry

Background:

  • A significant error in neutrophil identification by the IDEXX ProCyte Dx haematology analyser has been noted in feline blood samples.
  • This study investigated whether a similar neutrophil misclassification error occurs in canine blood samples using the same analyser.

Purpose of the Study:

  • To determine if the IDEXX ProCyte Dx analyser exhibits neutrophil misclassification errors in canine blood samples.
  • To assess the utility of white blood cell dot plot evaluation in detecting and preventing such erroneous results.

Main Methods:

  • Eighty-six canine blood samples were analyzed using the IDEXX ProCyte Dx, ADVIA 2120, and manual counting methods.
  • ProCyte Dx neutrophil counts were deemed unacceptable if they differed by more than 15% from both ADVIA 2120 and manual counts.
  • Instrument flags and white blood cell dot plots from the ProCyte Dx were critically reviewed.

Main Results:

  • The ProCyte Dx analyser yielded unacceptable neutrophil counts in 15% (13/86) of canine samples, showing counts significantly lower than reference methods.
  • Misclassified neutrophils were erroneously identified as monocytes and/or lymphocytes, particularly in samples from patients with systemic inflammation.
  • Reviewing ProCyte Dx white blood cell dot plots allowed for the identification and potential elimination of these erroneous results.

Conclusions:

  • The IDEXX ProCyte Dx analyser exhibits a common and severe neutrophil misclassification error in canine blood samples, potentially impacting clinical decisions.
  • Veterinary professionals operating the ProCyte Dx must meticulously evaluate white blood cell dot plots to ensure the accuracy of canine neutrophil counts and avoid diagnostic errors.