Evaluation of Sine Spin flat detector CT imaging compared with multidetector CT
Valentina D Petroulia1, Johannes Kaesmacher1, Eike I Piechowiak1
1Department for Diagnostic and Interventional Neuroradiology, Inselspital, Bern University Hospital and University of Bern, Bern, Switzerland.
Sine Spin Flat Detector Computed Tomography (SFDCT) shows high accuracy for detecting brain bleeds and early strokes. However, it has limitations in visualizing grey-white matter differentiation and infratentorial structures, with a reduced effective dose compared to multidetector CT.
Area of Science:
- Medical Imaging
- Radiology
- Neurological Imaging
Background:
- Flat detector computed tomography (FDCT) is utilized for periprocedural imaging in angiography.
- Sine Spin FDCT (SFDCT) employs a double oblique trajectory to minimize artifacts and enhance brain soft tissue imaging.
- This study compares the latest SFDCT generation with multidetector CT (MDCT) regarding dose, image quality, and diagnostic performance.
Purpose of the Study:
- To compare the effective dose, image quality, and diagnostic performance of Sine Spin FDCT (SFDCT) against multidetector CT (MDCT).
- To evaluate the diagnostic capabilities of SFDCT for detecting intracranial hemorrhage and early ischemic lesions.
Main Methods:
- Effective dose measurement using MOSFET detectors with the 7sDCT Sine Spin protocol on a biplane angiographic C-arm system.
- Diagnostic performance evaluation of SFDCT for anatomical details, hemorrhage, and ischemia detection, compared to MDCT.
- Calculation of inter- and intra-rater correlation, sensitivity, and specificity.
Main Results:
- SFDCT demonstrated equal diagnostic performance to MDCT in the supratentorial ventricular system.
- Image quality was inferior for SFDCT in grey-white matter differentiation and infratentorial structures.
- High sensitivity and specificity were observed for hemorrhage and early ischemic lesion detection with SFDCT; effective dose was 2 mSv.
Conclusions:
- SFDCT exhibits high diagnostic sensitivity and specificity for intracranial hemorrhage and early ischemic lesions.
- Limitations of SFDCT include delineation of grey-white matter and infratentorial structures.
- The 7sDCT Sine Spin protocol offers a lower effective dose compared to previous studies.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Magnetic Resonance Imaging