Electrophysiological Assessment of Murine Atria with High-Resolution Optical Mapping
Published on: February 22, 2018
On the Electrophysiology and Mapping of Intramural Arrhythmic Focus
Robert D Anderson1, Jairo Rodriguez Padilla2, Christian Joens1
1Hull Family Cardiac Fibrillation Management Laboratory, Division of Cardiology, University Health Network, Toronto General Hospital, Ontario, Canada (R.D.A., C.J., S.M., A.B., K.M., A.N., J.A., P.L., M.A.A., K.N.).
Background:
Conventional mapping of focal ventricular arrhythmias relies on unipolar electrogram characteristics and early local activation times. Deep intramural foci are common and associated with high recurrence rates following catheter-based radiofrequency ablation. We assessed the accuracy of unipolar morphological patterns and mapping surface indices to predict the site and depth of ventricular arrhythmogenic focal sources.
Methods:
An experimental beating-heart model used Langendorff-perfused, healthy swine hearts. A custom 56-pole electrode array catheter was positioned on the left ventricle. A plunge needle was placed perpendicular in the center of the grid to simulate arrhythmic foci at variable depths. Unipolar electrograms and local activation times were generated. Simulation models from 2 human hearts were also included with grids positioned simultaneously on the endocardium-epicardium from multiple left ventricular, septal, and outflow tract sites.
Results:
A unipolar Q or QS complex lacks specificity for superficial arrhythmic foci, as this morphology pattern occupies a large surface area and is the predominant pattern as intramural depth increases without developing a R component. There is progressive displacement from the arrhythmic focus to the surface exit as intramural focus depth increases. A shorter total activation time over the overlying electrode array, larger surface area within initial 20 ms activation, and a dual surface breakout pattern all indicate a deep focus.
Conclusions:
Displacement from the focal intramural origin to the exit site on the mapping surface could lead to erroneous lesion delivery strategies. Traditional unipolar electrogram features lack specificity to predict the intramural arrhythmic source; however, novel endocardial-epicardial mapping surface indices can be used to determine the depth of arrhythmic foci.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Electrophysiology of Normal Cardiac Rhythm
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...