Optimization of the Contrast Agent Injection Protocol for Carotid Artery Dynamic Contrast-Enhanced Magnetic Resonance
Yajie Wang1, Xiaoming Liu2,3, Yishi Wang4
1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China.
Background:
The injection protocol used in previous carotid artery dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) studies varied.
Purpose:
To investigate the effect of contrast injection protocol and optimize this protocol for carotid artery DCE-MRI.
Study Type:
Prospective.
Subjects:
Digital phantom and seven patients with carotid atherosclerosis.
Field Strength/Sequence:
3 T, spoiled gradient recalled echo sequence.
Assessment:
Different injection doses (0.01-0.3 mmol/kg) and effective injection rates (0.01-1 mmol/sec) were tested using a digital carotid plaque phantom considering the contrast pharmacokinetics, DCE-MRI imaging, contrast variation and flow-related imaging artifacts, random time delay between the contrast injection and image acquisition, and pharmacokinetic analysis process. For each injection protocol, combining the root mean square relative error (RMSRE) of the measured and maps within the adventitial vasa vasorum from 10 tested time delays by the root mean square produced RMSREoverall-vv which was used to measure the overall accuracy of the pharmacokinetic parameters. In vivo validation was performed on seven patients with carotid atherosclerosis by imaging them twice using the traditional commonly used protocol and the recommended protocol found by simulation.
Statistical Test:
Student's t-test, chi-square test, and paired t-test, P < 0.05 was considered statistically significant.
Results:
A low region of RMSREoverall-vv with the combination of medium injection dose and low effective injection rate was found. The protocol with injection dose of 0.07 mmol/kg and effective injection rate of 0.06 mmol/sec achieved the minimal RMSREoverall-vv (4.29%), thus was recommended, which showed more accurate arterial input function. Coinciding with the simulation results, this recommended protocol in in vivo experiments produced significantly fewer image artifacts, lower and (P all <0.05) than traditional protocol which overestimated these parameters in simulation.
Data Conclusion:
The contrast injection protocol influenced the accuracy of the pharmacokinetics parameter estimation in carotid artery DCE-MRI. The injection protocol with injection dose of 0.07 mmol/kg and effective injection rate of 0.06 mmol/sec was recommended.
Level Of Evidence:
2 TECHNICAL EFFICACY STAGE: 1.
More Related Videos
07:02Phase Contrast Magnetic Resonance Imaging in the Rat Common Carotid Artery
Published on: September 5, 2018
09:36A Magnetic Resonance Imaging-based Computational Protocol for Analysis of Plaque Morphology and Hemodynamics in Patients with Carotid Artery Stenosis
Published on: August 12, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies VII: Vascular Imaging
Magnetic Resonance Imaging