Related Experiment Video
Updated: Sep 29, 2025

Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
Effects of aggressive predilatation, sizing, and postdilatation strategy for coronary bioresorbable vascular
Hsiang-Yao Chen1,2,3, Wei-Chieh Huang1,2, Hsin-I Teng1,2,4
1Division of Cardiology, Department of Internal Medicine, Taipei Veterans General Hospital, Taipei, Taiwan, ROC.
Background:
The results of the recent Amsterdam Investigator-Initiated Absorb Strategy All-Comers trial showed that the predilatation, sizing, and postdilatation (PSP) technique did not lower the long-term rates of scaffold thrombosis and adverse events. We evaluated the impact of aggressive PSP bioresorbable vascular scaffold (BRS) implantation on the short- and long-term clinical outcomes.
Methods:
From June 2014 to December 2016, 150 patients with BRS implantation were enrolled and received successful percutaneous coronary intervention (PCI), of whom 104 received aggressive PSP technique (high-pressure predilatation and lesion preparation in addition to the traditional PSP technique). Short- and long-term outcomes were compared.
Results:
All patients underwent successful PCI and BRS implantation with final Thrombolysis in Myocardial Infarction grade 3 flow. The baseline and procedure characteristics were similar in both groups. Debulking techniques were used in 13 (8.7%) patients. Intracoronary imaging modalities were used in 73 (48.7%) patients. After BRS implantation, no adverse events were observed within 30 days in both groups. During the mean follow-up period of 2.98 ± 0.77 years, 12 (8.0%) patients experienced major adverse cardiovascular events (MACEs), including one cardiovascular death (0.6%), three nonfatal myocardial infarction (2.0 %), and 11 target-vessel revascularization (7.3 %). Multivariate Cox regression analysis showed that aggressive PSP remained an independent protective factor for MACEs. Moreover, the use of intracoronary imaging and rotablation atherectomy was associated with better clinical outcomes.
Conclusion:
Lesion preparation by aggressive PSP in BRS implantation was associated with better long-term clinical outcomes.