Shear wave elastography: the relationship of the cervical stiffness with gestational age and cervical length- a
Joyce Woo1, Weirong Ge2,1, Jyothi Mancheri1
1Sydney Institute for Women, Children and their Families, Sydney, Australia.
Objectives:
To standardize the measurement of shear wave elastography for assessment of cervical stiffness and its relationship with gestational age and cervical length.
Methods:
A prospective cross-sectional study was conducted from September 2017 to March 2019. Data from 125 unselected women (at 11-13 + 6, 18-22 and 24-28 weeks' gestation) and 55 high-risk women were analyzed for the study. Six regions of interest were evaluated for cervical elastography in the mid-sagittal position by transvaginal ultrasound. Statistical analyses were performed using R statistical language in R-studio. Delivery outcomes were recorded for each patient.
Results:
The shear wave elastography was feasible with good intraoperator and interoperator reproducibility. The endocervical canal and anterior lip internal position had the highest reproducibility (ICC-0.82, 0.75). Shear wave speed was significantly higher in all internal os regions than the external os. There was a statistically significant negative linear relationship of shear wave speed with the gestational age. There was a weak positive relationship between shear wave speed and cervical length. There was no difference between pregnancies with and without spontaneous preterm delivery in shear wave speed measurements and cervical length, although numbers were small for statistical analysis. The internal os of the large loop excision of the transformation zone group was stiffer than the normal population.
Conclusion:
Cervical elastography is feasible and effectively evaluates the tensile properties of the cervix during pregnancy. The most reproducible measurements were obtained at the anterior lip of the internal cervical os. Combining evaluation of cervical elasticity and length might further improve the identification of women at risk of preterm delivery. Currently, technical issues hinder the practical application of shear wave elastography in the clinical setting and require further research and development of the imaging modality.
More Related Videos
12:18Magnetic Resonance Elastography Methodology for the Evaluation of Tissue Engineered Construct Growth
Published on: February 9, 2012
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
Related Concept Videos
Elastic Strain Energy for Shearing Stresses
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity