First pointwise encoding time reduction with radial acquisition (PETRA) implementation for catheter detection in
Evangelia Kaza1, Casey Y Lee1, Martin T King1
1Brigham and Women's Hospital, Dana-Farber Cancer Institute, Harvard Medical School, Boston, MA.
Pointwise encoding time reduction with radial acquisition (PETRA) MRI successfully detected empty brachytherapy catheters with high accuracy. This imaging technique shows potential for MR-only treatment planning in interstitial HDR brachytherapy.
Area of Science:
- Medical Imaging
- Radiotherapy Physics
- Oncology
Background:
- Interstitial high-dose-rate (HDR) brachytherapy requires accurate catheter placement for effective treatment.
- Current imaging methods for catheter detection can have limitations in spatial accuracy and artifact susceptibility.
Purpose of the Study:
- To optimize and evaluate the Pointwise Encoding Time Reduction with Radial Acquisition (PETRA) sequence for detecting empty catheters in interstitial HDR brachytherapy.
- To assess the spatial accuracy and image quality of PETRA compared to conventional sequences.
- To evaluate the clinical feasibility of PETRA for MR-only treatment planning.
Main Methods:
- PETRA sequence was optimized and tested on phantoms containing empty catheters.
- Image quality and catheter detectability were assessed, comparing PETRA to T2-weighted sequences.
- Catheter tip position accuracy was evaluated using a treatment planning system (TPS).
- Clinical feasibility was assessed in a gynecological cancer patient with endometrial adenocarcinoma.
Main Results:
- PETRA provided positive signal for empty catheters with more accurate cross-sections than T2-weighted sequences (0.4 mm difference from nominal 2 mm diameter).
- PETRA exhibited minimal susceptibility differences and symmetric metal artifacts, unlike clinical sequences.
- PETRA-CT catheter tip position differences were < 1 mm, and interstitial catheter positional differences between PETRA and CT were < 1 mm on average.
- PETRA successfully detected interstitial catheters and templates in phantoms and a patient, fusing well with CT.
Conclusions:
- PETRA successfully and accurately detected interstitial catheters in phantoms.
- The first clinical implementation of PETRA demonstrates its potential for MR-only treatment planning in interstitial HDR brachytherapy.