Comparison of the oropharyngeal leak pressure between three second generation supraglottic airway devices during laparoscopic surgery in pediatric patients
Related Experiment Video
Updated: Sep 29, 2025

Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Abhyuday Kumar1, Chandni Sinha1, Neeraj Kumar2
1Department of Anesthesiology, AIIMS, Patna, India.
Background:
Previous studies have shown Proseal LMA and I gel similar to endotracheal intubation in ventilatory ability in pediatric laparoscopic surgeries.
Aims:
The primary aim of this study was to assess whether there is a significant difference in the oropharyngeal leak pressure between Ambu Auragrain, I-gel, and Proseal LMA during pediatric laparoscopic surgery.
Methods:
In this randomized controlled trial, 90 male patients of American Society of Anesthesiologists physical status I aged between 6 months and 10 years who were scheduled for laparoscopic single-sided inguinal hernia repair were recruited and randomly allocated to three groups in which airway was secured with Ambu Auragain, I gel, or Proseal LMA. The primary outcome was oropharyngeal leak pressure. The secondary outcomes were peak pressures before and after pneumoperitoneum, fiberoptic view, insertion attempts, insertion time, manipulations, perioperative and postoperative anesthesia-related problems. Continuous variables were compared using the one-way analysis of variance or the Kruskal-Wallis test with post hoc Turkey analysis. Categorical and ordinal data were compared using the chi-square test or Fisher's exact test.
Results:
Oropharyngeal leak pressure before pneumoperitoneum was higher with I gel as compared to Ambu Auragain (27.36 ± 5.72 cm of H2 O vs 23.56 ± 5.72 cm of H2 O) (p = .021) and PLMA (27.36 ± 5.72 cm of H2 O vs 23.24 ± 4.35 cm of H2 O) (p = .011). Oropharyngeal leak pressure after pneumoperitoneum was also higher with I gel as compared to Ambu Auragain (31.58 ± 4.35 cm of H2 O vs 26.83 ± 5.00 cm of H2 O) (p = .001) and Proseal LMA (31.58 ± 4.35 cm of H2 O vs 27.03 ± 3.80 cm of H2 O) (p = .002). Oropharyngeal leak pressures of Ambu Auragain and Proseal LMA were comparable. Postoperative complications were similar in all the supraglottic airway devices. No regurgitation or aspiration-related problem was observed in our study.
Conclusion:
I gel had a higher oropharyngeal leak pressure than the other two supraglottic airway devices and therefore may represent a better choice in situations where higher ventilatory pressures may be necessary, for example, in extremes of weight trendelenburg position, etc. CLINICAL TRIAL IDENTIFIER: Clinical trial registry of India (CTRI/2018/11/016445).
Related Concept Videos
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Suctioning the Nasopharyngeal Airway
Equipment Required
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...