Related Experiment Video
Updated: Sep 28, 2025

Cerebral Blood Oxygenation Measurement Based on Oxygen-dependent Quenching of Phosphorescence
Published on: May 4, 2011
Does cerebral oximetry always measure brain tissue oxygen saturation? An anatomical study utilizing computed
Rotem Naftalovich1,2, Darrick Chyu3, John T Denny4
1Medical Corps of the U.S. Army, U.S. Army Medical Department, 3630 Stanley Road, STE 104, Fort Sam Houston, TX, USA 78234.
Background And Aims:
to quantify the scalp-cortex distance and determine its variation among patients. We hypothesized that in a significant number of patients, this distance is greater than the maximum penetration depth of current cerebral oximeters.
Material And Methods:
A retrospective anatomic study using transverse head CT images selected randomly from 102 patients over the age of 18 years without brain swelling, intracranial mass effect, or brain hemorrhage. Scalp-cortex distances were determined at two separate locations along the craniocaudal axis; most cephalad to the frontal sinus (I0) and also 2 cm cephalad to that location (I2). Multiple measurements were obtained bilaterally at 1, 3, 5, 7, and 9 cm from midline.
Results:
The average scalp-cortex distance was 14.3 mm and 15 mm at I0 and I2 respectively. Distances varied more in I2 than in I0; from the measurements, 12.8% vs. 6.8% were over 20 mm, 4.4% vs. 2.2% over 25 mm, 1.1% vs. 0.6% over 35 mm and 0.6% vs. none over 40 mm at I2 and I0, respectively. 1.5% of the measurements at I2 were over 30 mm.
Conclusion:
Cerebral oximetry manufacturers all claim to measure cerebral tissue up to a depth of 20-25 mm; 20 mm with the EQUANOX and INVOS compared with 25 mm with the FORE-SIGHT. Scalp-cortex distance is within 25 mm in more than 95% of patients. However, even with the probe placed as per the manufacturer's recommendations, in a small but significant subset of patients, this distance is greater than the maximum penetration depth of current cerebral oximeters and hence may not reflect actual brain tissue oxygen saturation.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...