Assessment of postmanual cleaning adenosine triphosphate tests to prevent the use of contaminated duodenoscopes and
Arjan W Rauwers1, Anne F Voor In 't Holt2, Jolanda G Buijs3
1Department of Gastroenterology and Hepatology, Erasmus MC University Medical Centre, Rotterdam, The Netherlands; Department of Medical Microbiology and Infectious Diseases, Erasmus MC University Medical Centre, Rotterdam, The Netherlands.
Background And Aims:
We investigated whether the use of postmanual cleaning adenosine triphosphate (ATP) tests lowers the number of duodenoscopes and linear echoendoscopes (DLEs) contaminated with gut flora.
Methods:
In this single-center before-and-after study, DLEs were ATP tested after cleaning. During the control period, participants were blinded to ATP results: ATP-positive DLEs were not recleaned. During the intervention period, ATP-positive DLEs were recleaned. DLEs underwent microbiologic sampling after high-level disinfection (HLD) with participants blinded to culture results.
Results:
Using 15 endoscopes of 5 different DLE types, we included 909 procedures (52% duodenoscopes, 48% linear echoendoscopes). During the intervention period, the absolute rate of contamination with gut flora was higher (16% vs 21%). The main analysis showed that contamination was less likely to occur in the intervention period (odds ratio, .32; 95% credible interval [CI], .12-.85). A secondary analysis showed that this effect was based on 1 particular duodenoscope type (estimated probability, 39% [95% CI, 18%-64%] vs 9% [95% CI, 2%-21%]), whereas no effect was seen in the other 4 DLE types. In detail, of the 4 duodenoscopes of this type, 2 had lower contamination rates (69% vs 39% and 36% vs 10%). During the control period, both these duodenoscopes had multiple episodes with ongoing contamination with the same microorganism that ended weeks before the start of the intervention period (ie, they were not terminated by ATP testing).
Conclusions:
Postmanual cleaning ATP tests do not reduce post-HLD gut flora contamination rates of DLEs. Hence, postcleaning ATP tests are not suited as a means for quality control of endoscope reprocessing.
Related Concept Videos
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures V: ERCP
Patient...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Acute Kidney Injury IV: Diagnostic Studies and Prevention