Surgical Helmets Used During Total Joint Arthroplasty Harbor Common Pathogens: A Cautionary Note
Related Experiment Video
Updated: Sep 28, 2025

Novel Diagnostics in Revision Arthroplasty: Implant Sonication and Multiplex Polymerase Chain Reaction
Published on: December 3, 2017
Saad Tarabichi1, Emanuele Chisari1, Duncan S Van Nest1
1Rothman Orthopaedic Institute at Thomas Jefferson University, Philadelphia, PA.
Background:
The use of personal-protection surgical helmet/hood systems is now a part of the standard surgical attire during arthroplasty in North America. There are no protocols for the disinfection of these helmets.
Methods:
This is a prospective, single-center, observational study. Helmets worn by 44 members of the surgical team and foreheads of 44 corresponding surgical personnel were swabbed at three distinct time points. In addition, 16 helmets were treated with hypochlorite spray to determine if pathogens could be eliminated. Swabs obtained were processed for culture and next-generation sequencing (NGS).
Results:
Of the 132 helmet samples, 97 (73%) yielded bacteria on culture and 94 (71%) had evidence of bacterial-deoxyribonucleic acid (DNA) on NGS. Of the swabs sent for bacterial identification at the three time points, at least one from each helmet was positive for a pathogen(s). Of the 132 forehead samples, 124 (93%) yielded bacteria on culture and 103 (78%) had evidence of bacterial-DNA on NGS. The most commonly identified organism from helmets was Cutibacterium acnes (86/132) on NGS and Staphylococcus epidermidis (47/132) on culture. The most commonly identified organism from the foreheads of surgical personnel was Cutibacterium acnes (100/132) on NGS and Staphylococcus epidermidis (70/132) on culture. Sanitization of helmets was totally effective; no swabs taken the following morning for culture and NGS identified any bacteria.
Conclusion:
This study demonstrates that surgical helmets worn during orthopedic procedures are contaminated with common pathogens that can potentially cause surgical site infections. The findings of this study should at the minimum compel us to develop protocols for the disinfection of these helmets.
More Related Videos
04:37Author Spotlight: Advancing Research on Candida albicans Biofilm-Associated Prosthetic Joint Infections
Published on: February 2, 2024
08:30Surgical Retrieval, Isolation and In vitro Expansion of Human Anterior Cruciate Ligament-derived Cells for Tissue Engineering Applications
Published on: April 30, 2014
Related Concept Videos
Hand hygiene
Hand washing...
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Factors Affecting the Risk of Infection
The integrity and count of the white blood cells help the body resist pathogens and fight infection. When impaired, it reduces the body's resistance to pathogens. The acidic pH levels of the gastrointestinal, genitourinary tracts, and skin...
Healthcare Associated Infections I: Iatrogenic, Exogenic and Endogenic
HAIs significantly increase the cost of health care. Extended stays in healthcare institutions, increased disability, increased costs of medications, including specialized antibiotics, and prolonged recovery times add to the patient's expenses and the healthcare institution and funding bodies.