Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
The Impact of Smart Pump Interoperability on Errors in Intravenous Infusion Administrations: A Multihospital Before
Jilian Skog1, Sally Rafie2, Kumiko O Schnock3
1From the Sharp Healthcare, Chula Vista.
Objective:
The objective of this study was to assess the frequency, type, and severity of errors associated with intravenous medication administration before and after smart pump interoperability.
Methods:
We conducted an observational study at a community healthcare system before and after implementing smart pump interoperability. Point prevalence methodology was used to collect data on medication administration and errors in adult inpatient settings.
Results:
Observations were completed for 350 infusions preintervention (178 patients) and 367 postintervention (200 patients). Total errors significantly decreased from 401 (114.6 per 100 infusions) to 354 (96.5 per 100 infusions, P = 0.02). Administration errors decreased from 144 (41.1 per 100 infusions) to 119 (32.4 per 100 infusions, P = 0.12). Expired medication errors significantly reduced from 11 (3.1 per 100 infusions) to 2 (0.5 per 100 infusions, P = 0.02). Errors involving high-risk medications significantly reduced from 45 (12.8 per 100 infusions) to 25 (6.8 per 100 infusions, P = 0.01). Errors involving continuous medications significantly reduced from 44 (12.6 per 100 infusions) to 22 (6.0 per 100 infusions, P = 0.005). When comparing programming type, manual programming resulted in 115 (77.2%) of administration and user documentation errors compared with 34 errors (22.8%) that occurred when autoprogramming was used. Of these, errors involving high-risk medications reduced from 21 (84.0%) to 4 (16.0%) after using autoprogramming.
Conclusions:
Smart pump interoperability resulted in a 16% reduction in medication administration errors. Despite using dose error reduction software and autoprogramming, some types of errors persisted. Further studies are needed to understand how technology use can be optimized.
More Related Videos
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Errors occurring during blood pressure monitoring
Several factors...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Nonlinear Pharmacokinetics: Drug Elimination for IV Bolus Injection
Following the administration of a single intravenous (IV) bolus injection, we can determine the concentration of the drug in the plasma at any given time. This calculation is achieved using a specific equation that integrates the values of Vmax and KM.
We can also...