Radiolucent zones of biodegradable magnesium-based screws in children and adolescents-a radiographic analysis
Maximilian M Delsmann1,2, Julian Stürznickel1,2, Michael Kertai3
1Department of Trauma and Orthopaedic Surgery, Division of Orthopaedics, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Introduction:
Albeit the implantation of magnesium-based biodegradable implants can avoid a second surgery for implant removal, the postoperative occurrence of radiolucent zones around these implants based on corrosion processes has not been previously investigated in children and adolescents. We sought to characterize the distinct temporal and spatial dynamics for magnesium-based implants based on standard clinical routine radiographs.
Materials And Methods:
We retrospectively analyzed 29 patients, treated with magnesium-based compression screws (MAGNEZIX® CS 2.7 mm, CS 3.2 mm, CSC 4.8 mm; Syntellix AG) for fracture fixation, osteotomy, or osteochondral refixation. During a follow-up examination, the clinical and functional status was evaluated. Based on digital radiographs, the ratio of the area of the radiolucent zone and that of the screw was evaluated to assess implant degradation at two follow-up visits (i.e., after 6-8 weeks and 12-24 weeks).
Results:
In 29 patients (16/29 females, 14.03 ± 2.13 years), a total of 57 implants were evaluated that were used for osteotomy (n = 13, screws n = 26), fracture fixation (n = 9, screws n = 18), or osteochondral refixation (n = 7, screws n = 13). All patients healed without complications and regained full function. Radiolucent zones were observed in 27/29 patients at the first follow-up, with significantly decreased ratios at the second follow-up (2.10 ± 0.55 vs 1.64 ± 0.60, p = 0.0006). Regression analyses were performed to assess the temporal dynamics of radiolucent zones and revealed significant logarithmic developments for the 2.7 mm and 3.2 mm screws, marked by a strong ratio decrease during the first weeks and an almost complete disappearance after approximately 100 days and 200 days, respectively. In contrast, the ceramic-coated 4.8 mm screws presented a significant linear and slower decrease of radiolucent zones.
Conclusion:
Radiolucent zones are a common phenomenon in the course of implant degradation. However, they represent a self-limiting phenomenon. Remarkably, neither implant failure nor affected implant function is noted in this context. Yet, the microstructural changes accompanying the presence of radiolucent zones remain to be analyzed by three-dimensional high-resolution imaging.
Related Concept Videos
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...