Related Experiment Video
Updated: Sep 28, 2025

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Impact of 2 different posterior screw fixation techniques on primary stability in a cervical translational injury
Ludwig Oberkircher1, Julia Riemenschneider1, Martin Bäumlein1
1Center for Orthopaedics and Trauma Surgery, Philipps University of Marburg, University Hospital Giessen and Marburg GmbH, Location Marburg, Germany.
Background:
In case of injuries to the subaxial cervical spine, especially in osteoporotic bone, the question of the most stable operative technique arises. There are several techniques of screw fixation available regarding dorsal stabilization. This study investigates 2 techniques (lateral mass screws (LMS) vs cervical pedicle screws (CPS)) in the subaxial cervical spine regarding primary stability in a biomechanical testing using a translational injury model.
Methods:
A total of 10 human formalin fixed and 10 human fresh-frozen specimens (C 4 - T 1) were investigated. Specimens were randomized in 2 groups. Fracture generation of a luxation injury between C 5 and C 6 was created by a transection of all ligamentous structures as well as the intervertebral disc and a resection of the facet joints.Dorsal stabilization of C 4/C 5 to C 6/C 7 was performed in group A by lateral mass screws, in group B by pedicle screws. In the biomechanical testing, the specimens were loaded at 2 N/s in translation direction until implant failure.
Results:
Formalin fixed specimen: Mean load failure was 513.8 (±86.74) Newton (N) for group A (LMS) and 570.4 (±156.5) N for group B (CPS). There was no significant difference (P = .6905).Fresh frozen specimen: Mean load failure was 402.3 (±96.4) N for group A (LMS) and 500.7 (±190.3) N for group B (CPS). There was no significant difference (P = .4206).
Conclusion:
In our loading model respecting the translational injury pattern and a flexion movement we could not verify statistically significant differences between lateral mass screws and cervical pedicle screws. Mean loading failure was slightly higher in the CPS group though.