Related Experiment Video
Updated: Sep 28, 2025

A Structured Rehabilitation Protocol for Improved Multifunctional Prosthetic Control: A Case Study
Published on: November 6, 2015
Stacking classifier to improve the classification of shoulder motion in transhumeral amputees
1Electronics and Communication Department, Thapar Institute of Engineering and Technology, Patiala, Punjab 147001, India.
Abstract:
In recent years surface electromyography signals-based machine learning models are rapidly establishing. The efficacy of prosthetic arm growth for transhumeral amputees is aided by efficient classifiers. The paper aims to propose a stacking classifier-based classification system for sEMG shoulder movements. It presents the possibility of various shoulder motions classification of transhumeral amputees. To improve the system performance, adaptive threshold method and wavelet transformation have been applied for features extraction. Six different classifiers Support Vector Machines (SVM), Tree, Random Forest (RF), K-Nearest Neighbour (KNN), AdaBoost and Naïve Bayes (NB) are designed to extract the sEMG data classification accuracy. With cross-validation, the accuracy of RF, Tree and Ada Boost is 97%, 92% and 92% respectively. Stacking classifiers provides an accuracy as 99.4% after combining the best predicted multiple classifiers.