Published on: September 11, 2011
Joint replacement increases radiation exposure to the staff in angiography: A phantom study
A M Koenig1, A Akgöl1, J Verbe1
1Philipps-University of Marburg, Clinic of Diagnostic and Interventional Radiology, Marburg, Germany.
Introduction:
The present study investigates the influence of joint prostheses on the amount of scattered radiation in a simulated angiography set-up.
Materials And Methods:
A clinical angiography system (Artis Zee, Siemens Healthineers, Germany) with a water phantom as a scattering object was used. The scattered radiation of the water phantom was repeatedly measured without prosthesis, with a knee prosthesis and a hip prosthesis made from titanium, aluminum, vanadium, ceramics and polyethylene. For radiation measurement an ionization chamber (ambient dose equivalent rate from 0.1 µSv/h - 100 Sv/h, UMo, Berthold Technologies, Germany) was used. It was positioned on the right side of the phantom simulating an interventional procedure via the right femoral artery. The ionization chamber was positioned at 5 different heights (30, 100, 130, 150 and 165 cm), simulating different body parts of the interventionist. In addition, the amount of scattered radiation in relation to the tube angulation was investigated.
Results:
Averaged over all angulations at a height of 165 cm, the radiation dose was 2.7 times higher (1935 µSv/h, p < 0.01) when a hip prosthesis was present in comparison to no prosthesis (713 µSv/h). The radiation dose was 3.9 times higher with the integration of a knee prosthesis (2778 µSv/h, p < 0.01) compared to that without prosthesis. The average radiation dose over all angulations and all heights was 1491 µSv/h without prosthesis, 4538 µSv/h with a hip prothesis and 5023 µSv/h with a knee prosthesis respectively.
Conclusion:
This experimental study shows a significant increase in the radiation dose when a joint prosthesis is present in the examination field. Special attention and sufficient radiation protection is therefore necessary for investigations with implanted prostheses.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
X-ray Imaging