Accuracy of Potassium Measurement Using Blood Gas Analyzer
Hatim Mahmoud1, Zied Jaffar2, Yousef M Al Alawi1
1Diabetes and Endocrinology, Prince Mohammed Bin Abdulaziz Hospital, Al-Madinah, SAU.
Introduction:
Newer blood gas analyzers can measure both blood gases and electrolytes in both arterial and venous blood samples. They are small, compact, and mobile point of care test (POCT) devices. They can produce results in as short as five minutes. We aimed at assessing the accuracy of potassium (K) level measured by gas analyzer (index test) by comparing that to the regular laboratory machine (reference standard) in our hospital. Our goal is to use POCT result of potassium so we may start insulin infusion within five to 10 minutes of arrival of diabetic ketoacidosis (DKA) patients to the emergency room (ER). It takes an average of 30 minutes to get the result using the reference standard machine. Potassium level is needed urgently in cases of DKA before initiating insulin infusion. That is true also during cardiopulmonary resuscitation (CPR) and while replacing K in severe hypokalemia and during the management of hyperkalemia.
Methods:
We looked into the potassium results from 265 patients who had venous blood gas (VBG) or arterial blood gas (ABG) samples and compared that to results of potassium in venous blood samples of these same patients done simultaneously or within two hours. All patients who had blood gas and venous blood drawn simultaneously or within two hours were eligible irrespective of gender, age, diagnosis, and location in the hospital. Data were collected between January 2019 and June 2019. We excluded all cases that were receiving IV fluids, diuretics, or potassium supplements. Samples examined were from all different areas of the hospital including emergency room (ER), intensive care unit (ICU), and general floors. All ages and all diagnoses were included.
Results:
We used the Bland-Altman method to analyze our data. More than 95% of the data fell within ± 2 standard deviations (S) of the mean difference strongly suggestive of agreement between the index test and the standard reference of the laboratory methods. The bias was 0.19. Lin's concordance correlation coefficient was 0.6584.
Conclusion:
Findings of this study support the use of POCT blood gas analyzer for measuring potassium when the results are needed urgently. When measuring potassium, blood gas analyzers are as accurate as automated analyzers. They produce results in five minutes or so and can be relied upon when potassium level is needed urgently. They are cost-effective and may be available at the bedside.
Related Concept Videos
Flame Photometry: Lab
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Errors occurring during blood pressure monitoring
Several factors...
Disorders of Acid-Base Balance
Respiratory Acidosis and Alkalosis
Respiratory acidosis occurs due to an increase in the partial pressure of carbon dioxide PCO2 in the blood. It often arises from shallow breathing or impaired gas exchange caused by...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.