Lag screw osteosynthesis of simple olecranon fractures: A biomechanical comparative study
Ferdinand C Wagner1,2, Alexander Hamann2, Dirk Maier1
1Department of Orthopedics and Trauma Surgery, Medical Centre, Faculty of Medicine, Albert-Ludwigs-University of Freiburg, Freiburg, Germany.
Abstract:
Olecranon fractures are most frequently stabilized by tension band wiring (TBW), which unfortunately leads to relevant implant removal rates due to K-wire migration and soft tissue irritation. As lag screw osteosynthesis (LSO) might be a gentle and effective alternative in simple fracture patterns, the goal of the present study was to biomechanically compare LSO with TBW in simple olecranon fractures at a cadaver model. A simple olecranon fracture (Mayo type IIA) was created in eight pairs of human cadaver elbows, which were pairwise fixed by either TBW or two transcortical 4.0 mm lag screws. Biomechanical testing was conducted as a pulling force, applied to the triceps tendon in a 90° position. First, cyclic loading between 10 and 300 N was performed for 50,000 cycles. Afterward, maximum load was raised by 0.02 N/cycle until construct failure, what was defined as displacement >2 mm. Besides fracture displacement, failure cycle and failure load, the modes of failure were analyzed. Within the first five cycles, there was no significant difference in displacement (median TBW: 0.2 mm; LSO: 0.5 mm; p = 0.091). Both after 2000 (median TBW: 0.2 mm; LSO: 0.6 mm; p = 0.042) and after 20,000 cycles (median TBW: 0.4 mm; LSO: 0.9 mm; p = 0.027), the difference was significant. Failure cycle (median TBW: 72,639 cycles; LSO: 43,429 cycles; p = 0.017) and failure load (median TBW: 702 N; LSO: 303 N; p = 0.025) differed significantly as well. TBW mostly (6/8) failed at the lock of the cerclage wire, whereas most LSO constructs (5/8) failed as a pullout of the proximal fragment. In conclusion, to our biomechanical findings at human cadaver specimens, simple olecranon fractures treated by LSO show higher dislocation rates and lower failure loads compared to conventional TBW and mostly fail by pullout of the proximal fragment.
More Related Videos
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016