Diagnostic reference levels for chest computed tomography in children as a function of patient size
Denise Bos1, Sebastian Zensen2, Marcel K Opitz2
1Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, University of Duisburg-Essen, Hufelandstraße 55, 45147, Essen, Germany. denise.bos@uk-essen.de.
Background:
Radiation exposures from computed tomography (CT) in children are inadequately studied. Diagnostic reference levels (DRLs) can help optimise radiation doses.
Objective:
To determine local DRLs for paediatric chest CT performed mainly on modern dual-source, multi-slice CT scanners as a function of patient size.
Materials And Methods:
Five hundred thirty-eight chest CT scans in 345 children under 15 years (y) of age (median age: 8 y, interquartile range [IQR]: 4-13 y) performed on four different CT scanners (38% on third-generation and 43% on second-generation dual-source CT) between November 2013 and December 2020 were retrospectively analysed. Examinations were grouped by water-equivalent diameter as a measure of patient size. DRLs for volume CT dose index (CTDIvol) and dose-length product (DLP) were determined for six different patient sizes and compared to national and European DRLs.
Results:
The DRLs for CTDIvol and DLP are determined for each patient size group as a function of water-equivalent diameter as follows: (I) < 13 cm (n = 22; median: age 7 months): 0.4 mGy, 7 mGy·cm; (II) 13 cm to less than 17 cm (n = 151; median: age 3 y): 1.2 mGy, 25 mGy·cm; (III) 17 cm to less than 21 cm (n = 211; median: age 8 y): 1.7 mGy, 44 mGy·cm; (IV) 21 cm to less than 25 cm (n = 97; median: age 14 y): 3.0 mGy, 88 mGy·cm; (V) 25 cm to less than 29 cm (n = 42; median: age 14 y): 4.5 mGy, 135 mGy·cm; (VI) ≥ 29 cm (n = 15; median: age 14 y): 8.0 mGy, 241 mGy·cm. Compared with corresponding age and weight groups, our size-based DRLs for DLP are 54% to 71% lower than national and 23% to 85% lower than European DRLs.
Conclusion:
We developed DRLs for paediatric chest CT as a function of patient size with substantially lower values than national and European DRLs. Precise knowledge of size-based DRLs may assist other institutions in further dose optimisation in children.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography