Radiation dose optimization potential of deep learning-based reconstruction for multiphase hepatic CT: A clinical and
Yasunori Nagayama1, Makoto Goto2, Daisuke Sakabe2
1Department of Diagnostic Radiology, Graduate School of Medical Sciences, Kumamoto University, 1-1-1, Honjo, Chuo-ku, Kumamoto 860-8556, Japan.
Purpose:
This clinical and phantom study aimed to evaluate the impact of deep learning-based reconstruction (DLR) on image quality and its radiation dose optimization capability for multiphase hepatic CT relative to hybrid iterative reconstruction (HIR).
Methods:
Task-based image quality was assessed with a physical evaluation phantom; the high- and low-contrast detectability of HIR and DLR images were computed from the noise power spectrum and task-based transfer function at five different size-specific dose estimate (SSDE) values in the range 5.3 to 18.0-mGy. For the clinical study, images of 73 patients who had undergone multiphase hepatic CT under both standard-dose (STD) and lower-dose (LD) examination protocols within a time interval of about four-months on average, were retrospectively examined. STD images were reconstructed with HIR, while LD with HIR (LD-HIR) and DLR (LD-DLR). SSDE, quantitative image noise, and contrast-to-noise ratio (CNR) were compared between protocols. The noise magnitude, noise texture, streak artifact, image sharpness, interface smoothness, and overall image quality were subjectively rated by two independent radiologists.
Results:
In phantom study, the high- and low-contrast detectability of DLR images obtained at 5.3-mGy and 7.3-mGy, respectively, were slightly higher than those obtained with HIR at the STD protocol dose (18.0-mGy). In clinical study, LD-DLR yielded lower image noise, higher CNR, and higher subjective scores for all evaluation criteria than STD (all, p ≤ 0.05), despite having 52.8% lower SSDE (8.0 ± 2.5 vs. 16.8 ± 3.4-mGy).
Conclusions:
DLR improved the subjective and objective image quality of multiphase hepatic CT compared with HIR techniques, even at approximately half the radiation dose.