Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
Published on: December 9, 2010
Instrumental validation of free water, peak-width of skeletonized mean diffusivity, and white matter
Pauline Maillard1, Hanzhang Lu2, Konstantinos Arfanakis3,4
1Department of Neurology University of California, Davis Davis California USA.
Introduction:
To describe the protocol and findings of the instrumental validation of three imaging-based biomarker kits selected by the MarkVCID consortium: free water (FW) and peak width of skeletonized mean diffusivity (PSMD), both derived from diffusion tensor imaging (DTI), and white matter hyperintensity (WMH) volume derived from fluid attenuation inversion recovery and T1-weighted imaging.
Methods:
The instrumental validation of imaging-based biomarker kits included inter-rater reliability among participating sites, test-retest repeatability, and inter-scanner reproducibility across three types of magnetic resonance imaging (MRI) scanners using intra-class correlation coefficients (ICC).
Results:
The three biomarkers demonstrated excellent inter-rater reliability (ICC >0.94, P-values < .001), very high agreement between test and retest sessions (ICC >0.98, P-values < .001), and were extremely consistent across the three scanners (ICC >0.98, P-values < .001).
Discussion:
The three biomarker kits demonstrated very high inter-rater reliability, test-retest repeatability, and inter-scanner reproducibility, offering robust biomarkers suitable for future multi-site observational studies and clinical trials in the context of vascular cognitive impairment and dementia (VCID).