A retrospective analysis comparing set up errors from standard versus customised headrests for head and neck
E Anderson1, K Pilling1, S Iqbal1
1Northern Centre for Cancer Care Freeman Hospital, Freeman Road, Newcastle Upon Tyne, NE7 7DN, UK.
Radiography (London, England : 1995)
|April 10, 2022
Summary
Standard headrests reduce radiotherapy appointments but offer less reliable patient immobilization compared to customized headrests. Further review is needed for head and neck cancer radiotherapy services.
Area of Science:
- Radiation Oncology
- Medical Physics
- Cancer Treatment
Background:
- The COVID-19 pandemic necessitated reducing hospital visits for cancer patients undergoing radiotherapy.
- Standard headrests were implemented at the Northern Centre for Cancer Care (NCCC) as an alternative to customized headrests to minimize patient contact.
- Standard headrests require one mould room appointment, contrasting with three appointments for customized headrests.
Purpose of the Study:
- To compare the setup reproducibility and immobilization accuracy between standard headrests (SHR) and customized headrests (CHR) in head and neck radiotherapy.
- To evaluate the clinical implications of using standard headrests in terms of patient positioning during radiotherapy treatments.
Main Methods:
- A retrospective analysis of 20 patients (10 SHR, 10 CHR) treated between December 2019 and June 2020.
- Review of 547 cone beam computed tomography images to assess chin, shoulder, and spine positioning accuracy.
- Statistical analysis using the chi-squared test to compare positional deviations between the two headrest groups.
Main Results:
- Customized headrests demonstrated significantly better reproducibility for shoulder (p<0.001) and spine (p<0.001) positioning compared to standard headrests.
- Chin positioning was also more accurate with customized headrests (p=0.046).
- The customized headrest group had no setup re-scans, while the standard headrest group required 10 re-scans.
Conclusions:
- While standard headrests reduce patient hospital visits, potentially lowering COVID-19 exposure, customized headrests provide superior immobilization and setup reproducibility in head and neck radiotherapy.
- The findings indicate that customized headrests are more reliable for patient positioning accuracy.