Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024
Coronary Artery Radial Deformation and Velocity in Native and Stented Arteries
Logan S Schwarzman1, Decebal S Griza1, Leon J Frazin1
1Division of Cardiology, University of Illinois at Chicago, Chicago, IL, USA.
Introduction:
Coronary arteries are exposed to a variety of complex biomechanical forces during a normal cardiac cycle. These forces have the potential to contribute to coronary stent failure. Recent advances in stent design allow for the transmission of native pulsatile biomechanical forces in the stented vessel. However, there is a significant lack of evidence in a human model to measure vessel motion in native coronary arteries and stent conformability. Thus, we aimed to characterize and define coronary artery radial deformation and the effect of stent implantation on arterial deformation.
Materials And Methods:
Intravascular ultrasound (IVUS) pullback DICOM images were obtained from human coronary arteries using a coronary ultrasound catheter. Using two-dimensional speckle tracking, coronary artery radial deformation was defined as the inward and outward displacement (mm) and velocity (cm/s) of the arterial wall during the cardiac cycle. These deformation values were obtained in native and third-generation drug-eluting stented artery segments.
Results:
A total of 20 coronary artery segments were independently analyzed pre and poststent implantation for a total of 40 IVUS runs. Stent implantation impacted the degree of radial deformation and velocity. Mean radial deformation in native coronary arteries was 0.1230 mm ± 0.0522 mm compared to 0.0775 mm ± 0.0376 mm in stented vessels (p=0.0031). Mean radial velocity in native coronary arteries was 0.1194 cm/s ± 0.0535 cm/s compared to 0.0840 cm/s ± 0.0399 cm/s in stented vessels (p=0.0228).
Conclusion:
In this in vivo analysis of third-generation stents, stent implantation attenuates normal human coronary deformation during the cardiac cycle. The implications of these findings on stent failure and improved clinical outcomes require further investigation.