Low-rank inversion reconstruction for through-plane accelerated radial MR fingerprinting applied to relaxometry at
Nikolai J Mickevicius1, Carri K Glide-Hurst1
1Department of Human Oncology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Purpose:
To reduce scan time, methods to accelerate phase-encoded/non-Cartesian MR fingerprinting (MRF) acquisitions for variable density spiral acquisitions have recently been developed. These methods are not applicable to MRF acquisitions, wherein a single k-space spoke is acquired per frame. Therefore, we propose a low-rank inversion method to resolve MRF contrast dynamics from through-plane accelerated Cartesian/radial measurements applied to quantitative relaxation-time mapping on a 0.35T system.
Methods:
An algorithm was implemented to reconstruct through-plane aliased low-rank images describing the contrast dynamics occurring because of the transient-state MRF acquisition. T1 and T2 times from accelerated acquisitions were compared with those from unaccelerated linear reconstructions in a standardized system phantom and within in vivo brain and prostate experiments on a hybrid 0.35T MRI/linear accelerator.
Results:
No significant differences between T1 and T2 times for the accelerated reconstructions were observed compared to fully sampled acquisitions (p = 0.41 and p = 0.36, respectively). The mean absolute errors in T1 and T2 were 5.6% and 2.9%, respectively, between the full and accelerated acquisitions. The SDs in T1 and T2 decreased with the advanced accelerated reconstruction compared with the unaccelerated reconstruction (p = 0.02 and p = 0.03, respectively). The quality of the T1 and T2 maps generated with the proposed approach are comparable to those obtained using the unaccelerated data sets.
Conclusions:
Through-plane accelerated MRF with radial k-space coverage was demonstrated at a low field strength of 0.35 T. This method enabled 3D T1 and T2 mapping at 0.35 T with a 3-min scan.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Magnetic Resonance Imaging
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Atomic Nuclei: Types of Nuclear Relaxation
In spin–lattice or longitudinal relaxation, the excited spins exchange energy with the surrounding lattice as they return to the lower energy level. Among several mechanisms that contribute to spin–lattice relaxation, magnetic dipolar interactions are significant. Here, the excited nucleus transfers...