An Analytic Method for Calculating Scanner-, Kilovoltage Peak-, and Patient Size-Specific Hounsfield Unit Scale
Boyu Peng1, Sachin Jambawalikar1, Andrew J Einstein
1From the Departments of Radiology.
Objective:
This study aimed to calculate scanner-, kilovoltage peak (kVp)-, and patient size-specific computed tomography (CT) number thresholds for determining Agatston score (AgSc).
Methods:
The proposed method was validated using calcium measurements in an anthropomorphic phantom for 4 CT scanners made by 4 vendors. The derived mass concentration (γ) thresholds were used to calculate kVp- and patient size-specific CT number thresholds. Two models were applied to reduce intrascanner and interscanner AgSc variation, respectively.
Results:
The mean error of the modeled CT numbers is 1.8% (0.1%-4.4%). Model 1 has comparable results to the published phantom calibration method for an average-size patient (error, 1.5%; 0.1%-5.1%). The size- and the kVp-dependent fitting of modeled results have R2 greater than 0.965.
Conclusions:
Our results show a potential to enable accurate determination of AgSc under diverse conditions (eg, reduced tube potential) and are more easily applicable to different patient sizes than the phantom calibration method.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...