Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Prosthesis-Patient Mismatch in Small Aortic Annuli: Self-Expandable vs. Balloon-Expandable Transcatheter Aortic Valve
Jerome Ferrara1, Alexis Theron2, Alizee Porto2
1Département de Cardiologie, CHU Timone, 13005 Marseille, France.
Self-expandable (SE) transcatheter aortic valve replacement (TAVR) devices show better hemodynamic performance in patients with small aortic annuli compared to balloon-expandable (BE) devices, despite higher paravalvular leak rates. SE TAVR may be preferred for small aortic anatomies.
Area of Science:
- Cardiovascular Medicine
- Interventional Cardiology
- Biomedical Engineering
Background:
- Prosthesis-patient mismatch (PPM) negatively impacts outcomes after surgical aortic valve replacement (SAVR).
- PPM is increasingly recognized in transcatheter aortic valve replacement (TAVR), especially in patients with small aortic annuli.
- Evaluating hemodynamic performance of different TAVR devices in small annuli is crucial.
Purpose of the Study:
- To compare the hemodynamic performance of balloon-expandable (BE) Sapiens 3 and self-expandable (SE) Evolut Pro and Accurate Neo TAVR devices.
- To assess device performance specifically in patients with small aortic annuli (computed tomography aortic annulus area 330–440 mm²).
Main Methods:
- A prospective study enrolled 131 consecutive patients undergoing TAVR for small aortic annuli.
- Patients received either Sapiens 3 (BE), Evolut Pro (SE), or Accurate Neo (SE) devices.
- Hemodynamic performance, including prosthesis-patient mismatch (PPM) and paravalvular leaks (PVL), was evaluated.
Main Results:
- BE devices had significantly higher rates of PPM (39.5%) compared to SE devices (15.4% for Evolut Pro, 6.9% for Accurate Neo) (p < 0.0001).
- SE devices showed higher rates of moderate-to-severe paravalvular leaks (≥ 2/4) (15.4%–17.2%) compared to BE devices (2.6%) (p = 0.007).
- Despite higher PVL, SE devices achieved superior hemodynamic results.
Conclusions:
- Self-expandable TAVR devices demonstrate better hemodynamic performance in patients with small aortic annuli.
- The increased rate of paravalvular leaks with SE devices did not negate their hemodynamic advantage.
- SE TAVI devices are recommended as a potential first choice for patients with small aortic anatomy.