Related Experiment Video
Updated: Sep 27, 2025

Author Spotlight: Establishment and Confirmation of a Postnatal Right Ventricular Volume Overload Mouse Model
Published on: June 9, 2023
Quantification and Visualization of Reliable Hemodynamics Evaluation Based on Non-Contact Arteriovenous Fistula
Rumi Iwai1, Takunori Shimazaki2,3, Yoshifumi Kawakubo2
1Faculty of Clinical Engineering, Hyogo Medical University Hospital, Nishinomiya 663-8501, Japan.
Abstract:
The condition of arteriovenous fistula (AVF) blood flow is typically checked by using auscultation; however, auscultation should require a qualitative judgment dependent on the skills of doctors, and further attention to contact infection is required. For these reasons, this study developed a non-contact and non-invasive medical device to measure the pulse wave of AVFs by applying optical imaging technology. As a first step toward realization of the quantification judgment based on non-contact AVF measurement, we experimentally validated the developed system, whereby the hemodynamics of 168 subjects were visually and quantitatively evaluated based on clinical tests. Based on the evaluation results, the fundamental statistical characteristics of the non-contact measurement, including the average and median values, and distribution of measured signal-to-noise power ratio, were demonstrated. The clinical test results contributed to the future construction of quantified criteria for the AVF condition with the non-contact measurement.

