Lumped-Parameter and Finite Element Modeling of Heart Failure with Preserved Ejection Fraction
Published on: February 13, 2021
Computational Modeling of the Penn State Fontan Circulation Assist Device
Bryan C Good1, Sailahari V Ponnaluri1, William J Weiss1,2
1From the Department of Biomedical Engineering, Pennsylvania State University, University Park, Pennsylvania.
Abstract:
To address the increasing number of failing Fontan patients, Penn State University and the Penn State Hershey Medical Center are developing a centrifugal blood pump for long-term mechanical support. Computational fluid dynamics (CFD) modeling of the Penn State Fontan Circulatory Assist Device (FCAD) was performed to understand hemodynamics within the pump and its potential for hemolysis and thrombosis. CFD velocity and pressure results were first validated against experimental data and found to be within the standard deviations of the velocities and within 5% of the pressures. Further simulations performed with a human blood model found that most of the fluid domain was subjected to low shear stress (<50 Pa), with areas of highest stress around the rotor blade tips that increased with pump flow rate and rotor speed (138-178 Pa). However, the stresses compared well to previous CFD studies of commercial blood pumps and remained mostly below common thresholds of hemolysis and platelet activation. Additionally, few regions of low shear rate were observed within the FCAD, signifying minimal potential for platelet adhesion. These results further emphasize the FCAD's potential that has been observed previously in experimental and animal studies.
More Related Videos
06:12Modeling the Endothelial Glycocalyx Post-Pneumonectomy in a 3D Fluidic Chip - An Approach to Fabricating a Vascular-based Organ-on-Chip System
Published on: September 16, 2025
06:29Wireless Telemetry Device Implantation in a Fontan Ovine Model for Continuous and Long-Term Hemodynamic Monitoring
Published on: May 2, 2025