Defibrillator generator replacements in patients with left ventricular assist device support: The risks of hematoma
J J Eulert-Grehn1, I Sterner2, F Schoenrath1
1Department of Cardiothoracic and Vascular Surgery, Deutsches Herzzentrum Berlin (German Heart Center Berlin), Berlin, Germany; Deutsches Zentrum für Herz-Kreislauf-Forschung (German Center for Cardiovascular Research), partner site Berlin, Berlin, Germany.
Background:
The majority (89%) of left ventricular assist device (LVAD) patients have an implantable cardioverter-defibrillator (ICD) in place. Due to the advances of modern-day LVAD therapy, more patients are on support for longer. This inevitably leads to more LVAD patients facing ICD generator battery depletion. Until now, there are insufficient data regarding periprocedural risks of generator replacements in a high-risk group like the LVAD cohort.
Methods:
A retrospective, single-center analysis of pocket-related outcomes of all ICD generator replacements in LVAD and Non-LVAD patients between January 2014 and December 2018. The primary outcome was the combined endpoint of clinically significant pocket hematoma and/or cardiac implantable electronic device (CIED) infection in the first 6 months after ICD generator exchange. The clinically significant hematoma was defined as hematoma requiring reoperation, prolongation of hospitalization, or interruption of anticoagulation. The cumulative incidence function was calculated for the primary endpoint.
Results:
Two hundred seventy-seven patients underwent ICD generator exchange in our clinic in this time. Of these, 251 patients had a complete 6-month follow-up regarding clinically significant pocket hematomas and pocket infections. One hundred ninety patients had no LVAD, and 61 patients were on LVAD support. The rate of the primary combined endpoint clinically significant pocket hematoma and/or CIED infection was 3.5 times higher in LVAD patients compared to the non-LVAD cohort (event rate 39.14 vs 11.07 per 100 patient-years, p = 0.048). Clinically significant pocket hematomas necessitating revision occurred nearly 4 times more often in the LVAD group (p = 0.042). Pocket device infection rates were around 16 times higher in LVAD patients compared to non-LVAD patients (p = 0.002).
Conclusions:
Compared to Non-LVAD patients, LVAD patients exhibit a relevant higher rate of clinically significant pocket hematoma and CIED infection after ICD generator exchange. This information should additionally be considered in the decision-making process regarding the indication for ICD generator exchange.