Mass recalibration for desorption electrospray ionization mass spectrometry imaging using endogenous reference ions
Paolo Inglese1,2, Helen Xuexia Huang3, Vincen Wu3
1National Phenome Centre, Imperial College London, Hammersmith Campus, IRDB Building, London, W12 0NN, UK. p.inglese14@imperial.ac.uk.
Mass spectrometry imaging (MSI) data can suffer from mass calibration drift. This study introduces a computational method to correct mass errors in desorption electrospray ionization MSI, improving data reliability.
Area of Science:
- Analytical Chemistry
- Biomedical Imaging
Background:
- Mass spectrometry imaging (MSI) generates large datasets with thousands of mass spectra.
- Instrument calibration can drift during acquisition, leading to significant mass errors.
- These errors hinder data interpretation and affect spatial distribution analysis.
Purpose of the Study:
- To develop a post-acquisition computational method for mass recalibration in MSI.
- To address mass drift issues in desorption electrospray ionization MSI (DESI-MSI).
Main Methods:
- A computational approach utilizing known mass-to-charge ratios of common molecules.
- Testing on time-of-flight and Orbitrap mass analyzers coupled with DESI.
Main Results:
- Reduced observed mass drift by up to 60 ppm in biological samples.
- Improved molecular annotation accuracy.
- Enhanced spatial distribution quality of detected ions.
Conclusions:
- The method provides robust and accurate post-acquisition mass recalibration for DESI-MSI.
- Increases the reliability of molecular assignment and overall data quality.
More Related Videos
10:47Whole-body Mass Spectrometry Imaging by Infrared Matrix-assisted Laser Desorption Electrospray Ionization IR-MALDESI
Published on: March 24, 2016
09:38Dithranol as a Matrix for Matrix Assisted Laser Desorption/Ionization Imaging on a Fourier Transform Ion Cyclotron Resonance Mass Spectrometer
Published on: November 26, 2013
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometry: Overview
High-Resolution Mass Spectrometry (HRMS)
Matrix-Assisted Laser Desorption Ionization (MALDI)
The analyte of interest, a biomolecule or a mixture of biomolecules, is mixed with a suitable matrix material. The...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview