Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
A new visual quantitative assessment of ultrasound attenuation parameters for the mild liver steatosis
Xinping Ren1,2, Junqing Wang3, Shujun Xia1
1Department of Ultrasound, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Background:
Controlled attenuation parameter (CAP) without the guidance of the grey scale sonogram was a classic method in the quantitative evaluation of liver steatosis, it is recommended by international guidelines. Our study aimed to compare the diagnostic efficiency of a new real-time visual liver steatosis analysis (LiSA) versus CAP in chronic hepatitis B patients with liver steatosis.
Methods:
Patients were enrolled who underwent liver biopsy and received both LiSA (Hepatus, Mindray, probe LFP5-1U/s, China) and CAP (FibroScan502, Echosens, probe M, France) measurement simultaneously in our hospital from November 2018 to December 2019. The obtained values were both expressed as dB/m. Based on the liver fat content validated by liver biopsy, these patients were divided into the S0 group (fat content <5%) and S1 group (fat content ≥5%). The efficiency of the LiSA and CAP value in the diagnosis of liver steatosis was evaluated. Independent factors influencing the LiSA value were predicted by correlation analysis and multiple linear regression analysis.
Results:
A total of 151 patients were included in the analysis according to the exclusion criteria from 304 enrolled liver biopsy chronic hepatitis B (CHB) patients. Both LiSA and CAP successfully differentiated the S0 group from the S1 group. Receiver operating characteristic (ROC) curves showed that both LiSA and CAP had good diagnostic performance [area under the ROC curve area under the curve (AUC) >0.7] in evaluating liver steatosis, while there was no significant difference between the 2 methods (AUC 0.825 vs. 0.798, P=0.067). Using the optimal cutoff point, the specificity and sensitivity of LiSA in diagnosing liver steatosis were 89.18% and 79.16%, respectively. The specificity and sensitivity of CAP in diagnosing liver steatosis were 87.20% and 76.31%, respectively.
Conclusions:
Both LiSA and CAP are efficient for evaluating liver steatosis noninvasively.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
05:37Author Spotlight: A Non-Invasive Tool to Assess and Differentiate Fat Patterns in Liver Using 3D Dixon MRI
Published on: October 20, 2023
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Imaging Studies II: Ultrasonography