Performance of a Novel Handheld Bioelectrical Impedance Device for Assessing Muscle Mass in Older Inpatients
A K Stuck1, A Weber, R Wittwer
1Dr. med. Anna K. Stuck, Department of Geriatrics, Inselspital, Bern University Hospital, and University of Bern, Freiburgstrasse 46, CH-3010 Bern, Switzerland
The Journal of Frailty & Aging
|April 20, 2022
Summary
A handheld bioelectrical impedance analysis (BIA) device showed poor practicality and reliability for assessing muscle mass in older inpatients compared to a multisegmental BIA device. The handheld device had a high measurement failure rate, making it unsuitable for sarcopenia screening.
Area of Science:
- Geriatric medicine
- Biomedical engineering
- Clinical diagnostics
Background:
- Sarcopenia screening is crucial for older adults.
- Accurate body composition assessment is needed for effective sarcopenia management.
- Bioelectrical impedance analysis (BIA) offers a portable method for body composition analysis.
Purpose of the Study:
- To compare the practicality and repeatability of a handheld BIA device against a state-of-the-art multisegmental BIA device.
- To evaluate the suitability of a handheld BIA device for sarcopenia screening in older inpatients.
Main Methods:
- A cross-sectional study was conducted in a geriatric rehabilitation hospital.
- Appendicular skeletal muscle mass index (ASMI) was measured in 207 inpatients (aged 70+) using either a handheld or a multisegmental BIA device.
- Repeatability was assessed using intra-class correlation (ICC) and Bland-Altman plots.
Main Results:
- The handheld BIA device exhibited a significantly higher measurement failure rate (31%) compared to the multisegmental device (0.9%).
- Poor agreement and repeatability were observed with the handheld BIA device, indicated by Bland-Altman plots.
- The handheld device demonstrated lower practicality and reliability for assessing muscle mass in this population.
Conclusions:
- The handheld BIA device is not a practical or reliable tool for muscle mass assessment in older rehabilitation inpatients.
- Multisegmental BIA devices are superior for accurate sarcopenia screening in this demographic.
- Further research may be needed to refine handheld BIA technology for geriatric populations.
Keywords:
Appendicular skeletal muscle massbioelectrical impedance analysisgeriatricpracticalityrepeatabilitysarcopenia